INTERNATIONAL ROUND-ROBIN EXPERIMENT FOR SIMS-QUANTIFICATION
Abstract
Cross Calibration is an attempt to make relative sensitivity factors (RSFs) transferable between SIMS instruments and also usable by the same instrument at different times. The method relies on adjustment of carefully selected operating conditions and instrument tuning parameters, so that closely corresponding raw peak height ratios are obtained from the same sample by different instruments. The transferability of RSFs can be caracterised by the factors 1.68 and 2.40 for metallic glass and semiconductor matrices, respectively. Quantitative analysis of "unknown" samples resulted in an over al error factor of 1.31.
How to Cite
Riedel, M. “INTERNATIONAL ROUND-ROBIN EXPERIMENT FOR SIMS-QUANTIFICATION ”, Periodica Polytechnica Chemical Engineering, 34(1-3), pp. 73–80, 1990.
Issue
Section
Articles